LBW-NPS:面向纹理保真低剂量CT去噪的局部频带加权噪声功率谱损失
LBW-NPS: A Local Band-Weighted Noise Power Spectrum Loss for Texture-Preserving Low-Dose CT Denoising
研究提出局部频带加权噪声功率谱损失LBW-NPS,在重叠Hann窗patch内比较三个频带的谱能量,无需组织特异性损失条件。在DeepLesion模拟低剂量切片上,该方法将64×64像素评估窗内与损失相关的局部谱差距较池化频带加权NPS降低31.9%,PSNR无统计显著变化。均匀软组织中的平均NPS频率与GLCM误差同步下降,说明局部频带功率匹配可在不改动去噪架构的前提下改善区域纹理一致性。
arXiv:2611.25548v1 Announce Type: new Abstract: Low-dose CT denoising should suppress noise while retaining the local texture of the reference image. Matching a pooled spectrum can overlook regional discrepancies, whereas pixel losses alone can favor overly smooth outputs. We propose a local band-weighted noise power spectrum loss (LBW-NPS) that compares spectral energies within overlapping Hann-tapered patches, without tissue-specific loss conditioning. Each valid patch contributes a weighted squared log-energy difference across three frequency bands, alongside an $L_1$ reconstruction term. We compare LBW-NPS against $L_1$-only training and five auxiliary-loss baselines using the same U-Net and three seeds on DeepLesion slices with simulated low-dose acquisitions, including patch-scale ablations and lung, soft-tissue, and bone evaluations. On 220 test slices from 31 patients, LBW-NPS reduces a local spectral gap related to the loss, on $64\times64$-pixel evaluation windows, by 31.9% relative to pooled band-weighted NPS, without a statistically significant PSNR change. This improvement is supported by lower errors relative to the reference in average NPS frequency and gray-level co-occurrence matrix (GLCM) measures not used in the loss, evaluated in homogeneous soft tissue. Local band-power matching therefore improves regional texture agreement without changing the denoising architecture.
来源:arXiv (AI + Vision + Imaging) · arxiv.org