跳到正文
arXiv (Robotics + Hardware + Signal)· Douglas McDonough, Justin Liao, Jiawei Ge, Antony Goldenberg, Yvonne Zhang, Mackenzie A Eagleson, Ravi Prakash, Axel Krieger, Chunwoo Kim·· 7 天前AI 评分20

开源da Vinci X/Xi器械模组:面向腹腔镜手术研究的UR5/UR5e控制方案

An Open-Source da Vinci X/Xi Instrument Module for Laparoscopic Surgery Research

AI 导读

一套面向da Vinci X/Xi器械的开源模组可在UR5/UR5e机械臂上实现器械运动与电外科能量激活的同步控制与记录。电外科中间层优先响应术者脚踏板并记录每次激活请求来源,台架测试中完整记录了6次踩踏产生的全部12次激活与去激活事件。

正文 · 原文

arXiv:2611.21771v1 Announce Type: new Abstract: Studying electrosurgical tasks on a research robot requires control of both instrument motion and energy activation, along with a record of when each action occurs. This paper presents an open-source module for da Vinci X/Xi instruments on a UR5/UR5e robotic arm. Five motors transmit torque through timing belts to spring-loaded drive discs. An electrosurgical interposer accepts software commands, gives the surgeon's foot pedals priority, and records the source of each activation request. In bench testing, the interposer recorded all 12 activation and deactivation events from six pedal presses. Ex vivo demonstrations included monopolar and bipolar coagulation of splenic tissue and monopolar cutting during gallbladder dissection. Optical tracking of the Fenestrated Bipolar Forceps showed that calibration brought pitch and yaw gains from 0.904 and 1.14 to 1.00 and 1.02, respectively. The module enables researchers to record instrument motion, video, and the timing and source of electrosurgical commands during laparoscopic tasks.

来源:arXiv (Robotics + Hardware + Signal) · arxiv.org