arXiv (Robotics + Hardware + Signal)· Kavya Saravanan, Rohini Ramachandran, Jean Pierre Ndabakuranye, Isabelle Shearer, Gursimran Singh, Bobbi Fleiss, Omid Kavehei, Akram Al-Hourani, Francisco J. Tovar Lopez, Margaret Lech, Arman Ahnood·· 9 天前AI 评分36

无源无电池柔性超表面:商用60 GHz雷达实现植入式毫米波血糖监测

A passive, battery-free flexible metasurface for implantable millimetre-wave radar glucose sensing

AI 导读

一种无电子元件、无试剂的无源柔性超表面植入组织下方,由商用60 GHz雷达读取,用于追踪血糖水平。在生理盐水中,加入超表面后对数拟合R²从0.18提升至0.90;大鼠尸体模型在搏动血流下螺旋半径随葡萄糖浓度单调上升(R²=0.91),该单调响应可穿透最深6 mm组织。

正文 · 原文

arXiv:2611.17760v1 Announce Type: cross Abstract: Continuous glucose monitors (CGM) break the skin barrier and require periodic sensor replacement. We investigated whether a passive, battery-free flexible metasurface, implanted beneath tissue and read by a commercial 60 GHz radar, can track blood glucose levels. The metasurface contained no electronics or reagents. We fitted a conical helix to the in-phase and quadrature radar signals across the frequency sweep and used its mean radius (the spiral radius) as the glucose-dependent feature. We tested glucose concentrations of 25 to 400 mg/dL in saline solution with and without the metasurface, and 196 to 596 mg/dL of total glucose in porcine whole blood beneath ex vivo porcine tissue and in a murine cadaver model with pulsatile blood flow. The metasurface was essential for achieving a monotonic sensor response; in saline, the coefficient of determination ($R^2$) of a logarithmic fit increased from 0.18 without the metasurface to 0.90 with it. In a murine (rat) cadaver model, the signal's spiral radius increased monotonically with glucose concentration despite pulsatile flow ($R^2$ = 0.91), and this monotonic behavior persisted through up to 6 mm of tissue. Although sensitivity decreased at higher glucose levels, each discrete concentration step remained clearly detectable. These results demonstrate that a shallow, passive implant paired with a commercial radar can effectively track blood glucose without requiring onboard electronics or batteries.

来源:arXiv (Robotics + Hardware + Signal) · arxiv.org