10.5T人脑MRI局部屏蔽16通道发射阵列的电磁/电路联合仿真与并行发射优化
Optimization of a Locally Shielded 16-Channel Transmit Array for Human Brain MRI at 10.5 T via Electromagnetic/Circuit Co-Simulation and Parallel Transmission Strategies
研究通过电磁/电路联合仿真结合并行发射策略,优化了用于10.5T人脑神经成像的局部屏蔽16通道发射RF阵列。筛选出的对齐环形折叠偶极子(LFD)阵列配紧凑圆柱屏蔽(半径160mm、窗口长70mm),在梯度屏蔽集成后无需RF电路重调,外部场泄漏低于1%,全脑激发系数变异在静态RF匀场下约27%、动态pTx下约9%。未屏蔽LFD阵列实测B1+效率分布与仿真吻合,NRMSE约8%。
arXiv:2611.34395v1 Announce Type: new Abstract: Modern neuroimaging and neuroscience increasingly demand high spatial and temporal resolution, driving the use of ultrahigh-field (7 T and above) MRI for improved signal-to-noise ratio. Realizing this advantage depends critically on advanced radiofrequency (RF) and gradient technologies, particularly multi-channel transmit arrays and high-performance head-gradient systems. In integrated systems, a local RF shield must provide sufficient field confinement while preserving transmit performance and minimizing sensitivity to nearby conductive structures. In this work, we sought to optimize a locally shielded 16-channel transmit RF array for human neuroimaging at 10.5 T using electromagnetic/circuit co-simulation combined with parallel-transmission strategies. Candidate transmit-array and RF-shield configurations were evaluated before and after gradient-shield integration to identify a configuration with suitable RF transmit and shielding performance for the integrated system. The results show that no single standalone metric, including static RF shimming performance, dynamic parallel transmission (pTx) performance, or external-field leakage, reliably predicts robustness after gradient-shield integration. Progressive screening identified the aligned loop-folded dipole (LFD) array combined with a compact cylindrical shield featuring a conical top and three windows (160-mm radius and 70-mm window length) as the final configuration. Without RF-circuit retuning, the selected design maintained external-field leakage below 1% and after gradient shield integration achieved whole-brain excitation coefficient of variation (CoV) of approximately 27% with static RF shimming and 9% with dynamic pTx. Experimental validation of the unshielded LFD array showed good agreement between simulated and measured B1+ efficiency distributions, with a normalized root-mean-square error (NRMSE) of approximately 8%.
来源:arXiv (Optics + Display + Materials) · arxiv.org