研究团队报道了一种紧凑型统一量子电学平台,可在约4 K的单一低温恒温器中同步实现量子电压与量子电阻,并依欧姆定律导出量子电流。分层磁屏蔽在270 mm轴向间距内实现6 T与低于50 nT共存,约瑟夫森与量子霍尔子系统分别以2.6E-9和1.4E-8的相对标准不确定度符合量子化预期值。经改进的低温电流比较器链接后,实现相对不确定度6.6E-8的50 μA量子电流。

正文 · 原文

A co-located integration quantum electrical standard is essential to reduce reliance on distributed traceability in high-accuracy metrology, especially for portable and on-site use. Metrologically, realizing any two of voltage, resistance, and current is sufficient, as the third follows from Ohm's law. The combination of Josephson voltage and quantum Hall resistance offers better uncertainty, but conflicts with the tesla-level field for quantum Hall and near-zero field for Josephson operation. Here we report a compact unified platform enabling co-realization of quantum voltage and resistance in a single cryostat near 4 K, with quantum current derived via Ohm's law. A hierarchical magnetic shielding with staged attenuation and spatial confinement allows 6 T and below 50 nT to coexist within 270 mm axial separation with negligible cross-coupling. In integrated operation, the Josephson and quantum Hall subsystems agree with expected quantized values within relative standard uncertainties of 2.6E-9 and 1.4E-8, respectively. Linking them via an improved cryogenic current comparator realizes a 50 μA quantum current with relative uncertainty of 6.6E-8. These results demonstrate that three basic electrical units can be synchronously realized with superior metrological consistency on a single integrated platform, offering a viable transition from distributed calibration chains toward compact-integrated quantum-based realization.

来源:arXiv (Instruments + Metrology) · arxiv.org