arXiv (Optics + Display + Materials)· Daniel Tandeitnik, Gabriel Dias, Thiago Guerreiro·· 12 天前AI 评分18

基于FPGA实时外差解调的悬浮纳米颗粒绝对位移校准

Absolute displacement calibration of a levitated nanoparticle with real-time heterodyne demodulation

AI 导读

研究团队提出一种基于FPGA数字同相/正交解调的外差方案,用于光悬浮纳米颗粒等光机械系统的位置读出,无需锁相、信号与颗粒位移呈线性,且校准不受本振与散射场光功率漂移影响。

正文 · 原文

arXiv:2607.09912v2 Announce Type: replace-cross Abstract: We report a heterodyne scheme for the position readout of an optomechanical system, in particular an optically levitated nanoparticle, based on digital in-phase and quadrature demodulation on a field-programmable gate array. Compared to the standard homodyne approach, it offers three key advantages: it requires no phase lock, it produces a signal linearly proportional to the particle displacement, and its calibration is immune to drifts in the optical power of the local oscillator and of the scattered field. Combined with a vectorial model of the focal field, it measures the displacement in length units without free parameters. Through simultaneous homodyne and heterodyne measurements on the same trapped particle we demonstrate these advantages: the heterodyne follows a driven particle linearly up to amplitudes at which the homodyne response folds back, and calibrates the homodyne, whose standard thermometric calibration we find biased by 33 %. Inverting equipartition, the calibrated readout measures the mass of the particle with an uncertainty below 1 %, after removing a bias of up to 20 % caused by absorption heating at 1550 nm. The method can be used in any optomechanical system based on phase readout.

来源:arXiv (Optics + Display + Materials) · arxiv.org